Calibrating the axes of the swing arm profilometer by the four-based laser trackers
Gao, Mingxing1,2; Jing, Hongwei1; Fan, Tianquan1; Chen, Lin1,2; Li, Jie1
Volume9623
Pages96230M
2015
Language英语
ISSN0277-786X
DOI10.1117/12.2186278
Indexed BySCI ; Ei
Subtype会议论文
AbstractThe swing arm profilometer (SAP) realizes the measurement of the different surface shape by adjusting the axis of the arm bearing (AAB) and the axis of the worktable bearing (AWB) for different spatial positional relationships, so the positional relationship that has the high-accuracy calibration between the AAB and the AWB is the premise to obtain accurate result. This paper presents a way using the four-based laser trackers to calibrate the axes of the SAP ,that satisfies the requirement of the high precision and the dynamic calibration. The four-based laser trackers system only uses the precise distances to obtain the coordinates of points. The axes can be fitted by the coordinates of points in the rotate arcs, and the relationship between the axes can be known. The study of the measurement error caused by the deviation of the calibration ,the study of the self-calibration of the four-based laser trackers and the simulation of the measurement are included in this paper. © 2015 SPIE.; The swing arm profilometer (SAP) realizes the measurement of the different surface shape by adjusting the axis of the arm bearing (AAB) and the axis of the worktable bearing (AWB) for different spatial positional relationships, so the positional relationship that has the high-accuracy calibration between the AAB and the AWB is the premise to obtain accurate result. This paper presents a way using the four-based laser trackers to calibrate the axes of the SAP ,that satisfies the requirement of the high precision and the dynamic calibration. The four-based laser trackers system only uses the precise distances to obtain the coordinates of points. The axes can be fitted by the coordinates of points in the rotate arcs, and the relationship between the axes can be known. The study of the measurement error caused by the deviation of the calibration ,the study of the self-calibration of the four-based laser trackers and the simulation of the measurement are included in this paper. © 2015 SPIE.
Conference NameProceedings of SPIE - The International Society for Optical Engineering
Conference Date2015
Citation statistics
Document Type会议论文
Identifierhttp://ir.ioe.ac.cn/handle/181551/7909
Collection质量处(质检中心)
Affiliation1. Institute of Optics and Electronics, Chinese Academy of Science, Chengdu, Sichuan, China
2. University of Chinese Academy of Sciences, Beijing, China
Recommended Citation
GB/T 7714
Gao, Mingxing,Jing, Hongwei,Fan, Tianquan,et al. Calibrating the axes of the swing arm profilometer by the four-based laser trackers[C],2015:96230M.
Files in This Item:
File Name/Size DocType Version Access License
2015-2158.pdf(618KB)会议论文 开放获取CC BY-NC-SAApplication Full Text
Related Services
Recommend this item
Bookmark
Usage statistics
Export to Endnote
Google Scholar
Similar articles in Google Scholar
[Gao, Mingxing]'s Articles
[Jing, Hongwei]'s Articles
[Fan, Tianquan]'s Articles
Baidu academic
Similar articles in Baidu academic
[Gao, Mingxing]'s Articles
[Jing, Hongwei]'s Articles
[Fan, Tianquan]'s Articles
Bing Scholar
Similar articles in Bing Scholar
[Gao, Mingxing]'s Articles
[Jing, Hongwei]'s Articles
[Fan, Tianquan]'s Articles
Terms of Use
No data!
Social Bookmark/Share
All comments (0)
No comment.
 

Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.